Specifications |
879 |
878A |
Parameters Measured |
L/C/R/D/Q/_ |
L/C/R/D/Q |
Measuring Circuit Mode |
Inductance (L)?Defaults to series mode |
Capacitance/ Resistance (C/R)-Defaults to parallel mode |
Displays |
L/C/R: Maximum display 19999 D/Q: Maximum display 999 (Auto Range). |
Ranging Mode |
Auto & Manual |
Measuring Terminals |
3 terminals with sockets |
Test Frequency
Accuracy: ±0.1 % |
100Hz=100 Hz |
120Hz= 120 Hz |
120Hz= 120 Hz |
1KHz =1010 Hz |
1KHz =1010 Hz |
|
10KHz= 9.6 KHz |
|
Backlit display |
Included |
Excluded |
Tolerance mode |
1%, 5%, 10%, 20% |
1%, 5%, 10% |
Test Signal Level |
0.6Vrms approx. |
Measuring Rate |
1 time/second, nominal |
Response time |
Approx. 1 second/ DUT (device under test)
(@ manual range) |
Auto Power-Off |
5 minutes approx. without operation |
Temperature Coefficient |
0.15 x (Specified Accuracy) / °(0-18° or 28-40°) |
Operation Temperature |
0° to 40°; 0-70% R.H. |
Storage Temperature |
-20° to +50°; 0-80% R.H. |
Low Battery Indication |
Approx. 6.8V |
Power Consumption |
Approx. 40mA for operation/ 0.08mA after Auto Power-off. |
Power Requirements |
1) DC 9V Battery
2) Ext. DC Adaptor: DC 12Vmin ?15Vmax. (Load 50mA Min.) |
Protective Fuse |
0.1A/250V Fuse (input protective) |
Standard Accessories |
Test alligator clips (pair) DC 9V Battery. User manual |